2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 330
CO- 1.0.8 [CONDUCTIVITY] MORE THAN 14.5 mS / ISOLATE PUMP ELECTRICAL CONNECTIONS
A) These procedures are IMPORTANT to prevent patient harm.
BE CAREFUL!
B)
Figure BELOW
, touch the top of the
ACID
AND
BICARB
pumps
to feel them
BOTH
mechanically stroking about every three (3) seconds!
C)
IMPORTANT!
Return the BICARB (blue) connector to its rinse port!
D) Call debug screen 2 and ENSURE
BICOUT
= 0!
E) Touch the
ACID AND BICARB PUMPS
again. TWO (2) possible scenarios:
1)
IF (and ONLY if) the ACID PUMP continues to stroke BUT the BICARB PUMP stops:
This
is normal! See procedure number
(page 331).
2)
IF the BICARB PUMP is stroking AND the ACID PUMP stops:
See parts a AND b below:
a)
IMPORTANT! To prevent damage turn the machine OFF!
b) Figure above, the acid (“Conc-P”) and bicarbonate pump (“Bic-P”) distribution board
connectors are reversed!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...