2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 394
CO- 11.0.00 RETURN SYSTEMS
a) Return Pre Dialyzer Temp Sensor NTC #3s connector to
2
nd
distribution board position from the LEFT, “MON-
NTC”.
b) From debug screen 0, ENSURE Pre Dialyzer Temp reads
more than 35.0° C.
c) If removed from a previous procedure, return Pre Dialyzer
Cond Cell #7 connector to distribution board position “x7,
COND”.
d) From debug screen 0, ENSURE Pre Dialyzer Cond reads more than 10.0 mS.
e) The following components (see
COMPONENT LIST
below) are to be swapped in, with known good,
one at a time then, in between, continue to parts f through i to see if the new component fixes the
problem.
COMPONENT LIST:
1)
Power Logic Board;
2)
Sensor Board
1
;
3)
Sensor Board cable;
4)
Functional Board
2
;
5)
Distribution board.
1
To prevent “Cond Offset Failure” place the machine into T and C mode. Refer to
OPERATING
MODES
, (page
Error! Bookmark not defined.
).
2
To prevent “Cond Offset Failure” the machine MUST AGAIN be placed into T and C mode!
f) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
g) From the Home screen, set [Dialysate Flow] to 500 ml/min and press ‘CONFIRM’!
h) Allow six (6) FULL minutes, for Temperature and Conductivity to stabilize to normal, before continuing
to part i!
i)
Read before performing!
Repeat (ABOVE) procedure number
page 378
) to see if the
new component fixed the problem but if you return here, continue through the Component List until the
OLC problem is fixed.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...