2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 500
Figure 84 – Hydraulics Rear View
DAIR- 4.0.0 ISOLATE VALVE #39
If Valve #39 is sticking open target Deaeration Pressure CANNOT be achieved:
a) Per the Figures above AND right locate
Valve #39 on the BOTTOM of the
hydrochamber.
b) Figure right, clamp Valve #39’s clear
tubing at the location seen.
c) Can
[PUMP RATE]
now be adjusted to
achieve
target pressures?
Yes
Target pressures achieved! See procedure number
(page 501).
No Pressures CANNOT be achieved! See procedure number
page 502
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...