2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 655
DVF- 2.0.0 VERR = 0 / ISOLATE VALVE #24 OPEN CIRCUIT
Call debug screen 2 to see
DIAVLO
(Figure below). THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) DIALVLO REMAINS ALWAYS = 1:
Electrical problem with Valve #24. Proceed to
page 657
, procedure number
2)
IF DIAVLO REMAINS ALWAYS = 0:
With the shunt door REMAINING open, see procedure number
3)
IF (and ONLY if) DIAVLO blinks rapidly between 0 and 1:
Electrical problem with Valve #24.
Proceed to
page 657
, procedure number