2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 237
f) Figure below, plug in the 10 pin ribbon cable to Functional Board’s P8 connector.
g)
Close the card cage!
h) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
i)
Allow the machine to run for several hours to see if a temperature problem
reoccurs:
1)
IF a temperature problem DOES NOT reoccur:
A temperature problem is
not occurring at this time. DO NOT continue!
2) IF a temperature problem reoccurs:
The BTM MAY be bad. Check its dip
switch settings per its Operator’s Manual.
LEFT BLANK INTENTIONALLY