2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 491
AIR- 1.0.1 CHECK FOR AN AIR LEAK AT FILTER #73
Figure below, for two (2) minutes, watch through the OUTPUT (to machine) side of Dialysate Filter #73.
This is
Location #1 (L1)
. Air bubbles seen?
Yes
Air seen at L1! See procedure number
(page 492).
No air seen at L1! Proceed to
page 493
, procedure number
Figure 74 – Leak Checks (1) / Abbreviated Flow Diagram
Figure 75 – Hydraulics (External & Rear Views)