2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 589
DIASAFE- 6.0.0 ISOLATE DIASAFE
®
TEST VALVE #28
a) An
air-filled
60 ml syringe is REQUIRED!
b) ENSURE the syringe plunger moves freely back and forth inside the barrel!
c) Move the syringe plunger to the 40 ml mark.
d) Figure below, attach the syringe to Valve #28’s tubing.
e)
Reset ALL alarms!
f) Repeat the DiaSafe
®
Test while watching the syringe for
one (1) minute
. Is the plunger drawn into
the barrel?
Yes
Plunger drawn! Valve #28 is OKAY! Proceed to
page 591
, procedure number
No
Plunger is NOT drawn! See procedure number
(page 589).
DIASAFE- 6.1.0 PLUNGER IS NOT DRAWN IN / ISOLATE FOR VALVE #28 SOLENOID ACTIVATION
Per the Figure above, touch Valve #28’s (black) solenoid. Is it warm?
Yes
Solenoid is warm! TWO (2) possible bad components:
1)
Bad Actuator-Test Board OR
2)
Bad
Valve #28.
No Solenoid IS NOT warm! Perform parts a THROUGH d next page: