2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 591
DIASAFE- 7.0.0 VALVE #28 OKAY
a) Return FILTER #92 to Valve #28!
DIASAFE- 8.0.0 DIASAFE TUBING ATTACHED PROPERLY
a) This procedure uses a psi pressure gauge.
ENSURE
it
reads 0 psi before installing it!
b) Figure right, tee the gauge between the Flow
Pump's OUTPUT nozzle and its WHITE tubing.
c) Clamp both sides of the gauge tubing to prevent
leaks and false readings!
d)
IMPORTANT! Place the machine into
RINSE!
e) Watch for one (1) minute to ENSURE
a “No Water” alarm
NEVER occurs!
f) Call debug screen 0. If the debug screens do not appear press the ‘Esc’ key then call screen 0.
g) Allow Valve #43’s ‘dot’ (Figure right) to turn blue then WHITE again! While
white, does pressure CYCLE, about every three (3) seconds, to between 35
and 36 psi?
Yes
Between 35 and 36 psi! See procedure number
(page 592).
No
Is NOT between 35 and 36 psi! ENSURING the machine was in RINSE AND no leaks,
TWO (2) possible scenarios:
1)
IF (and ONLY if) pressure is too low:
DO NOT calibrate instead proceed to
page 93
, procedure number
2)
IF pressure is too high:
Perform parts a AND b below:
a) Per the Figure (above, right) adjust Valve #78 until pressure cycles to between 35
and 36 psi!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...