2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 454
TMP- 2.0.8 ISOLATE ACCURATE DIALYSATE FLOW
a)
IMPORTANT!
Press the “1” key
for five (5) seconds.
b)
Allow forty five (45) seconds,
BEFORE continuing to part
c!
c) Call debug screen 5.
QdS
(
Figure
right)
REMAINS = 1 if dialysate
flow is stable.
d) Watch
QdS
for two (2) minutes
OR until if it EVER = 0. TWO (2)
possible scenarios?
1)
IF (and ONLY if) QdS REMAINS = 1:
(page 454).
2) IF QdS EVER = 0!
See parts a AND b below:
a) Press the “1” key for five (5) seconds THEN allow forty five (45) FULL seconds before
continuing to part b!
b) Return to debug screen 5. Does
QdS
NOW REMAIN = 1?
Yes
QdS
REMAINNS = 1! See procedure number
(page 454).
No
QdS
goes to 0 at least once! Call debug screen 0 to watch
Flow Error
for three
(3) FULL minutes. If it EVER = 1 proceed to
page 20
. If NEVER = 1 a restriction is indicated. Try
replacing the DiaSafe
®
filter, rinse the machine for five minutes, then return to
(page 452). If you return here there may
a restriction to the drain.
TMP- 2.0.9 QDL REMAINING BETWEEN 720 AND 880
Call the Home screen. Is the TMP window REMAINING white?
Yes
TMP window white! Proceed to
page 456
, procedure number
No TMP window RED! See parts a AND b below:
a) Attempt the ALARM RESET SEQUENCE up to twice BEFORE continuing to part b.
b) Allow thirty (30) seconds! What color is the TMP window now? TWO (2) possible scenarios
next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...