2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 546
OVER- 5.0.0 ISOLATE VALVE #39 (Slow Overflow)
Valve #39 opens in Cleaning Programs and Deaeration pressure should be NOWHERE NEAR -24 inHg! This
is especially important for Heat Disinfect to prevent boiling which would cause overflow!
a)
IMPORTANT!
Re-attach Valve #41’s the ‘air gap tower’ tubing.
b) The deaeration gauge is used next. ENSURE it reads 0 inHg before installing it!
c) Per the Figure below, tee the gauge between the inlet (clear tubing) and the deaeration pump’s input
nozzle.
Figure 102 – Deaeration Pump Head
d) Return the concentrate connectors to their rinse ports
e)
IMPORTANT! Place the machine into RINSE!
TWO (2) possible scenarios:
1)
IF (and ONLY if) pressure is between 0 and negative (-)12 inHg.
Valve #39 is okay! See
(page 547).
2)
IF pressure is between negative (-)18 and negative (-)25 inHg:
Valve #39 is NOT opening! FOUR
(4) possible bad components:
1)
Bad Actuator-Test Board OR;
2)
3)
Bad ACTUATOR cable OR;
4)
Bad distribution board!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...