2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 195
TROUBLESHOOTING VALVE ERRORS IN DIALYSIS PROGRAM
If
VERR
= one (1) or more and / or is continuing to increase indicates a valve circuit resistance of less than
20 Ω and is used for troubleshooting. HOWEVER, if
VERR
reaches it maximum of 255 it becomes useless!
A) From here forward, Figure right, if (and ONLY if) a “Dial
Valve Failure” OR “Act Byp Valve Fail” banner EVER
appear proceed to
page 652
B)
CLOSE the SHUNT door!
C) ENSURE a “No Water” banner NEVER appears!
D) FIGURE BELOW, a previous procedure may have already placed a resistor plug into the distribution
board’s AIR SENSOR position
4
th
connector cap i.e. 5
th
position
FROM THE LEFT!
E) TWO (2) possible scenarios below:
1)
IF (and ONLY if) a resistor plug is ALREADY in place:
(page 196).
2)
IF a resistor plug IS NOT in place:
See parts a THROUGH e next page:
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...