2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 559
LEAKING- 3.0.0 LEAK IN CLEANING PROGRAM / ISOLATE SPECIAL VALVE ERROR
a) Press ’CONFIRM’ to return to the Main Service Mode menu.
b) Select the screen’s ‘Options’ button.
c) Call debug screen 0. IGNORING the top window (Flow Error) look at
Valve
Error
(2
nd
window down)
. TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve Error = 1 LONGER THAN two (2) seconds:
Proceed
page 189
2) IF Valve Error does NOT = 1 for LONGER THAN two (2) seconds:
Was the leak ORIGINALLY
caused because UF Check Valve tubing was ‘blowing off’?
Yes
UF Check Valve tubing was ‘blowing off’! See procedure number
No UF Check Valves were okay! Turn the machine OFF then proceed to
page 560
, procedure
LEAKING- 3.1.0 LEAK IN CLEANING PROGRAM / UF CHECK VALVE TUBING BLOWING OFF
a) Turn the machine off then back on!
b) Subsequent squirting into the bucket occurs if (and ONLY if) UF Check Valve #63 is okay!
c) If the screen’s Heat Disinfect button is BLUE go to Heat Disinfect. If GRAY go to
RINSE
!
d) Figure right, if (and ONLY if) no leaks from the Fluid Sample port = UF Check
Valve #64 okay.
e) Call debug screen 0 to watch
Valve Error
(2
nd
window down)
until if it =1
for LONGER THAN two (2) seconds OR until an active leak reoccurs OR for
three (3) FULL minutes, whichever comes first. TWO (2) possible scenarios
below:
1)
IF (and ONLY if) Valve Error EVER = 1 for LONGER THAN two (2) seconds:
Proceed page
, procedure number
2) IF Valve Error does NOT = 1 for longer than two (2) seconds:
THREE (3) possible scenarios i) or
ii) or iii) below:
i)
IF (and ONLY if) an active leak reoccurs:
Turn the machine OFF then see procedure
(page 560).
ii) IF (and ONLY if) an active leak does NOT reoccur HOWEVER, a Flow Error occurs:
Proceed to
page 158
, procedure number
(page 158).
iii)
IF an active leaks does NOT reoccur AND a Flow Error NEVER occurs:
Reconnect the
Fluid Sample If UF tubing blows off again one or both of the Check Valves may be bad!
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...