2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 229
T- 1.0.155 ISOLATE TEMPERATURE MONITOR
a) Figure right, plug the
6.04 KΩ
resistor
plug, from the
TWO-RESISTOR
SET
into
2
nd
distribution board position
from the left, “MON-NTC”!
b) Call debug screen 4.
c) Watching
TEMP
(lower left) for one (1) minute, THREE (3) possible scenarios 1) or 2) or 3) below:
1)
IF (and ONLY if) between 4.0 and 6.0 AND STABLE i.e. does NOT change more than 0.2:
See
(page 229).
2) IF (and ONLY if) less than 4.0 OR more than 6.0:
Perform
parts a AND b below:
a) Remove the plug then reinsert it into the
2
nd
position from the left, “MON-NTC”. If
TEMP
is
still less than 4.0 OR more than 6.0 see part b. If now between 4.0 and 6.0 AND STABLE
(not changing more than 0.2 per minute) see procedure number
(page 229).
b) Use the
6.04 KΩ
plug from a different
TWO-RESISTOR SET
. If (and ONLY if)
TEMP
is still
less than 4.0 OR more than 6.0 proceed to
page 295
, procedure number
. If
TEMP
between 4.0 and 6.0 AND STABLE (not changing more than 0.2 per minute) see procedure
number
(page 229).
3)
IF between 1.0 AND 8.0 BUT changes more than 0.2 per minute:
Perform parts A
through C below:
A) Leaving the plug in place, turn the machine OFF.
B) FOUR (4) possible bad components, see COMPONENT LIST below. Swap in each, one at a
time, and in between continue to part C) to test the new component! COMPONENT LIST:
1)
Sensor Board
*
;
2)
Functional Board*;
3)
Power Logic Board;
4)
Distribution board
*
For each board, to prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
))
C) Return to Dialysis Program. Watch debug screen 4’s
TEMP
for one (1) minute. If now between
4.0 and 6.0 and stable the last component swapped in is bad. If still not between 4.0 and 6.0 or
unstable return to part A to try the next component in the list!
T- 1.0.16 TEMP STABLE BETWEEN 4.0 AND 6.0 / VERIFY TEMPERATURE
Call the Home screen. Based on the [
Temperature
] window, TWO (2) possible scenarios:
1)
IF (and ONLY if) between 35.1 and 38.9 °C:
ENSURING the [
Temperature
] window is pale
yellow/white, see procedure number
(page 230).
2) IF less than 35.1 OR more than 38.9 °C:
Proceed to
page 302
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...