2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 25
Chamber #69 says “Air’ continued:
d) Use a different plug from the
four-resistor
set! If (and ONLY if) the text box now says “No
Air” always see
(page 26). If it EVER says “Air” continue to part e.
e) The text box continues to say “Air” OR toggles between “Air” and “No Air”. SIX (6) possible bad
components:
1)
CBE board
1
;
2)
Sensor Board
2
;
3)
Sensor cable
3
4)
Functional Board
4
;
5
) Distribution board;
6)
Motherboard.
1
A)
Swap in a known good CBE board;
B)
Return to Dialysis Program (“Select Program” →
‘Dialysis’ → ‘CONFIRM’);
C)
Plug one of the resistors, from the four-resistor set, into the
new CBE board;
D)
From debug screen 0, if the text box now says “No Air” always the
previous CBE board is bad.
2
Leaving the resistor plug installed:
A)
With the machine off, swap in a known
good Sensor Board;
B)
To prevent “Cond Offset Failure” place the machine into
T and C
Mode
(refer to
OPERATING MODES
(page
Error! Bookmark not defined.
);
C)
Return to
Dialysis Program);
D)
From debug screen 0, if the text box now says “No Air” always the
previous Sensor Board is bad.
3
Leaving the resistor plug installed:
The Sensor Board cable can be checked.
NOTE
that one (1)
AIR SENSOR
connection will be checked and proceed to
page 524
SECTION 17 - CHECKING THE SENSOR BOARD CABLE
4
Leaving the resistor plug installed:
A)
With the machine off, swap in a known good
Functional Board;
B)
To prevent “Cond Offset Failure” place the new Functional Board into
T
and C Mode
;
C)
Return to Dialysis Program;
D)
From debug screen 0, if the text box now
says “No Air” always the previous Functional Board is bad.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...