2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 391
CO- 10.0.60 ISOLATE POST SENSORS STABILITY
a) Per the Figure below, remove Post Dialyzer Temp Sensor NTC #44’s connector from distribution board
position “x44, NTC-POST”.
b) Using a flashlight, check the vacant “x44” position for ‘white’ corrosion and/or damaged
male
pins.
If
damage is located this may be the problem!
c) Per the Figure right, place the
6.04
KΩ
plug, from the
TWO-RESISTOR SET
into the Post Dialyzer Temp Sensor’s vacant distribution board position “x44,
NTC-POST”.
d) Per the Figure above, remove Post Dialyzer Cond Cell #13’s connector from
distribution board position “x13, COND-POS”.
e) Check the vacant “x13” position for ‘white’ corrosion and/or damaged
male
pins.
f) Place the
274 Ω
resistor plug, from the
TWO-RESISTOR SET
, into the Post Dialyzer Cond Cell’s #13
distribution board position “x13, COND-POS”.
g) Call debug screen 5. TWO (2) checks then proceed according to the TWO (2) possible scenarios after
check #2:
CHECK #1 (TPOS):
WITHOUT LOOKING AWAY, watch
TPOS
(4
th
column from
right, at top) for exactly one (1) minute noting its highest and lowest values. Subtract
the lowest value seen from the highest. 2 (two) or less = GOOD! MORE THAN 2 =
BAD!
CHECK #2 (CPOS):
Now watch
CPOS
(4
th
column from right, third box down) for
exactly one (1) minute noting its highest and lowest values. Subtract the lowest
value seen from the highest. 8 (eight) or less = GOOD! MORE THAN 8 is BAD!
1)
IF (and ONLY if) TPOS AND / OR CPOS ARE BAD:
Proceed to
page 394
procedure number
2)
IF TPOS AND CPOS are both GOOD:
See parts a through d below:
a) Return Post Dialyzer Temp Sensor NTC #44’s connector to distribution board position labeled
“x44, NTC-POST”.
Parts b through d next page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...