2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 451
TMP- 2.0.4 FILACT REMAINING = 0
a) Figure right, from here forward, if (and ONLY if) a “Dial
Valve Failure” OR “Act Byp Valve Fail” banner EVER
appears proceed to
page 652
b)
IMPORTANT! Turn the machine OFF to prevent damage!
c)
Figure right
, at the TOP of the
distribution board, ENSURE the Sensor
ribbon cable is plugged in SECURELY.
d)
Figure right
, the Air Sensors’s position (5
th
position from the left) MUST BE VACANT!
e) Leaving the position VACANT, turn the machine
on.
f) Return to Dialysis Program
(“Select Program →
‘Dialysis’ → ‘Enter’)!
g) Allow three (3) minutes BEFORE continuing to part
h!
h) Call debug screen 1 to see
FILACT
. TWO (2) possible scenarios:
1)
IF (and ONLY if) FILACT now = 1: A)
IMPORTANT!
Return the female Air Sensor
PROPERLY to its distribution board position;
B)
(page 452).
2) IF FILACT REMAINS = 0:
Perform
parts a AND b below:
a) BEFORE continuing to part b, NOTE this page number then perform
b) Leaving the Air Sensor’s female connector unplugged until
FILACT
= 1, FIVE (5) possible bad
components:
1)
CBE board;
2)
Sensor Board
a
;
3)
Functional Board
b
;
4)
Distribution board;
5)
Motherboard.
a
A)
Swap in a known good Sensor Board (see
B)
Place the machine into
T and C
Mode
(refer to
OPERATING MODES
, page
Error! Bookmark not defined.
);
C)
IMPORTANT!
Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ’CONFIRM’);
D)
From debug screen 1.
if
FILACT
= 1 now the previous Sensor Board is bad.
b
A)
Swap in a known good Functional Board (see
B)
Place the new
Functional Board into
T and C Mode;
C)
Return to Dialysis Program;
D)
From debug screen 1, if
FILACT
= 1 now the previous Functional Board is bad.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...