2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 633
P- 1.0.9 X2 / PIN 3 LESS THAN 6.0 VOLTS / ISOLATE MAIN RELAY (K1)
a) Turn the machine off and
UNPLUG
it.
CAUTION!
Electrocution hazard if not unplugged!
b) Slide the power supply away from the cabinet to see
the Power Control board inside.
c) ENSURE the cables, along the top edge of the board,
are plugged in securely.
If not this may be the
problem!
d) Un-mount the board from its four plastic clips.
e) Figure right, remove the two (2)
front
screws and lay
power supply rear panel down to access to the rear
(solder) side of the board.
f)
IMPORTANT!
Plug the machine in.
CAUTION! High voltage now present!
Figure 124
g)
Per the Figure below
, measure from the rear side of the Power Control board’s four-pin X3
connector at
pin 1
. More than 6.0 volts DC?
Yes
More than 6.0 volts! TWO (2) possible bad components: 1) Most likely the twenty-pin Power
Logic cable (X2) OR; 2) Main Relay K1 OR possibly its four-pin cable
1
.
No Less than 6.0 volts! Unplug the machine then TWO (2) possible bad components:
1)
Most
likely, bad main Relay K1 (possibly it four-pin cable)
1
OR
2)
Bad Power Control board.
1
To LOCATE K1 and its cable refer to
(page 651).
Figure 125 – Power Control Board / X3 Connector, Pin 1
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...