2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 138
MOTORS- 5.0.0 ISOLATE INTERMITTENT MOTOR
a)
IMPORTANT!
Turn the machine OFF!
b) Figure right, remove the pump head from the motor to
expose the ‘Drive Magnet’.
c) Manually spin the ‘Drive Magnet’. Does it rotate freely i.e. is
NOT ‘binding’ (Yes or No)?
Yes
Rotates freely!
See procedure number
No
Does NOT rotate freely! The motor is bad
*
and brushes won’t help.
*
NOTE:
The binding motor may have destroyed the Actuator-Test and/or Power Logic Board
MOTORS- 5.0.2 MOTOR ROTATES FREELY
a) ENSURE rotating the drive magnet causes the motor shaft to rotate.
b) If the shaft rotates a problem is not indicated at this time HOWEVER, the motor may be stalling
intermittently. Replace the brushes but if the motor stops again in the near future, THREE (3) possible
bad components below:
COMPONENT LIST:
1)
Bad Motor OR;
2)
Bad pump head OR;
3)
Bad Actuator-Test Board*. Swap in each one at a time
with known good and test for a stalling motor in between.
* To LOCATE the board refer to
(page 9)
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...