2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 339
CO- 2.0.7 ‘ACTUAL’ CONDUCTIVITY BETWEEN THE ALARM LIMITS
a)
IMPORTANT!
Press ‘CONFIRM’!
b) Figure right, with ‘Actual’ Conductivity REMAINING
BETWEEN the Limits, what color is the Conductivity
window now?
1)
IF (and ONLY if) WHITE:
See procedure
page 340
).
2)
IF RED:
Call debug screen 3 to see
HICOND
AND
LOCOND
(lower right). TWO (2) possible scenarios:
Scenario #1: IF (and ONLY if) BOTH HICOND AND LOCOND = 0!
Select
the ‘Dialysate’ tab then see procedure number
Scenario #2:
IF HICOND = 1 OR LOCOND = 1!
THREE (3) possible bad
components:
1)
Actuator-Test Board; 2) Functional Board;
3)
Motherboard.
CO- 2.0.7.1 BOTH HICOND AND LOCOND = 0 AND THE CONDUCTIVITY WINDOW IS RED
Allowing up to three (3) minutes, does the [Conductivity] window eventually turn white?
Yes
[Conductivity] window white! See procedure number
(page 340).
No After three (3) minutes the [Conductivity] window STAYS RED! Perform parts a THROUGH d
below:
a) If a previous procedure placed
the 274Ω resistor plug into the Conductivity Cell’s
distribution board position remove it and return Conductivity Cell #7’s connector to
distribution board position “X7, COND”.
b) Close the shunt door and return the concentrate connectors to their rinse ports.
c)
IMPORTANT! Place the machine into RINSE!
d) Is the external flow indicator’s ‘bob’ rising at least ¼ way up in the sight tube?
Yes
‘Bob’ moving! See procedure number
(page 353).
No Bob NOT moving! See parts a THROUGH c below:
a) Plug into acid and bicarb jugs.
b) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
c) Proceed to
page 20
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...