2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 353
CO- 6.0.0 DISTRIBUTION BOARD CHECKS / ON LINE CLEARANCE (OLC) ENABLED?
a) Figure right,
inside the distribution board, TWO
checks:
CHECK #1:
ENSURE position #4, “PH-PR” is
VACANT!
CHECK #2:
ENSURE Conductivity Cell #7 AND
NTC #3’s connectors are plugged in
properly.
b) If troubleshooting an Online Clearance (OLC) problem see part c. If NOT, proceed to
page 354
c) Call debug screen 0.
d) Figure right, do the
Post Dialyzer Cond
and
Temp
symbols
appear (Yes or No)?
Yes
The OLC option is enabled! See procedure
number
No The OLC option is off!
A)
Enter Service mode → Options → Hardware Options;
B)
Set OLC to
“Yes” and press ‘CONFIRM’;
C)
Return to debug screen 0 to ENSURE the
Post Dialyzer
Cond
and
Temp
symbols now appear;
D)
Proceed to
page 354
, procedure number
CO- 6.1.0 OLC OPTION ENABLED
a) Figure right, ENSURE Conductivity Cell #13’s
AND NTC #44’s connectors are plugged in
properly.
b) Open the shunt door.
c) Allow thirty (30) seconds.
d) Is screen 0’s
Post Dialyzer Temp
more than
34.00° C?
Yes
Post Dialyzer Temp
is more than 34.00° C! See procedure number
(page 354).
No
Post Dialyzer Temp
LESS THAN 34.00° C! Proceed to
page 380
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...