2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 346
CO- 4.2.0 ISOLATE CONCENTRATE INLET SYSTEMS (2)
.
a) Return the acid and bicarb CONNECTORS
HARD
into their rinse ports.
b)
IMPORTANT!
Place the machine into RINSE!
c) ENSURE no external leaks!
(page 346).
CO- 4.2.1 ISOLATE PUMP FEEDBACK
a) Call debug screen 1. Watch
NO EOS
AND
ALWEOS
(third column from right) for
one (1) FULL minute.
Table 7 – EOS
NO EOS
ALWEOS
YOUR RESPONSE
0
(always)
0
(always)
(page 346).
Does not matter
1
(EVER)
Proceed to
page 397
1
(EVER)
Does not matter
Proceed to
page 397
CO- 4.2.2 ISOLATE POTENTIAL LEAKS
a) Figure right, ENSURE NO external LEAKS
at the Acid AND Bicarb Pumps.
b) Plug into acid and bicarb!
c) Return to Dialysis Program (“Select
P
rogram” → ‘Dialysis’ → ‘CONFIRM’)!
d) Allow forty-five (45) seconds BEFORE
continuing to part e.
e) Is air / concentrate now drawn,
TOWARDS
THE MACHINE, in ONE direction ONLY
?
Yes
Air / concentrate now drawn into the machine in ONE DIRECTION ONLY! See procedure
number
No Air / concentrate NOT seen OR moving back and forth! Proceed to
page 348
, procedure
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...