2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 148
NW- 2.0.0 VALVE #41’S ‘DOT ALWAYS BLUE = CONSTANT “NO WATER” ALARM
a) Figure below, ENSURE Valve #41’s connector is plugged PROPERLY into distribution board position
“V27, IN-V”.
b) Call debug screen 0 to watch Valve #41’s ’dot’ for one (1) FULL minute. TWO (2) possible scenarios:
1)
IF (and ONLY if) Valve #41’s ‘dot’ is now remaining white OR cycling between white and
blue:
The constant “No Water” alarm is no longer occurring! If (and ONLY if) a “No Water” alarm
occurs intermittently return to (ABOVE) procedure number
page 146
). If a “No
Water” alarms NEVER occurs the No Water problem is not occurring!
2) IF Valve #41’s ‘dot’ is still NEVER white i.e. ALWAYS blue:
Ignoring
Flow Error
, look at the
2
nd
window down, Valve
Error!
TWO (2) possible scenarios:
Scenario #1: IF (and ONLY if) Valve Error = 0!
(page 149).
Scenario #2: IF Valve Error EVER = 1 longer than two (2) seconds!
Refer to the Table
below:
If the machine is CURRENTLY in:
Your response
Dialysis Program i.e. connected to
acid and bicarbonate
Proceed to
page 652
A Cleaning / Disinfection Program
(Rinse, Heat Disinfect, etc.)
Proceed to
page 189
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...