2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 344
Figure 57 – Hydraulics Rear View
CO- 3.4.0 [CONDUCTIVITY] WITHIN 0.4 OF TCD / VERIFY CONDUCTIVITY ALARM LIMITS
What color is the Home screen’s [
Conductivity
] window now? TWO (2) possible
scenarios:
1)
IF (and ONLY if) pale yellow / white:
2) IF RED:
See parts a THROUGH e below:
a) At the bottom of the screen, select the
‘Dialysate’ tab.
b) Figure right, using [Alarm Position], adjust the
Limits so the ‘Actual’
Conductivity
is
CENTERED between them.
c) Press ‘CONFIRM’!
d) Allow up to two (2) additional minutes for the ‘Actual’ Conductivity window to turn white!
(page 344).
CO- 3.5.0 VERIFY MEASURED CONDUCTIVITY
a) Allowing up to two (2) additional minutes OR until the external flow indicator’s ‘bob’ starts to moving up
and down!
b) Based on the meter’s reading versus the machine’s [
Conductivity
] window. TWO (2) possible
scenarios:
1)
IF (and ONLY if) the meter and the [Conductivity] window are within 0.15 of each other:
Proceed to
page 375
, procedure number
2) IF the meter and the [Conductivity] window are NOT within 0.15 of each other!
Proceed to
page 353
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...