
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 331
CO- 1.0.8.1 [CONDUCTIVITY] MORE THAN 14.5 mS / ISOLATE PUMP TUBING CONNECTIONS
a) Per the top Figure below, trace the tubing from the red (ACID) connector to ENSURE it goes to the
input of the
ACID PUMP
b) If ABSOLUTELY SURE the pump is plumbed correctly, see procedure number
(page 332).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...