2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 590
a) Trace Valve #28’s wires from the
distribution board to the valve’s
(black) solenoid. THREE (3)
checks:
Check #1:
If a wire is pinched or
insulation is damaged this
may be the problem!
Check #2:
If the wires DO NOT terminate properly at the solenoid terminals this may be the
problem!
Check #3:
If the solenoid terminals are corroded this may be the problem!
b) Figure right, open Valve #28’s female
distribution board connector cap.
c) If the wires ARE NOT connected between the
TOP and BOTTOM terminals this is the
problem!
d)
Measure resistance (Ω), between the TOP and
BOTTOM
terminals. Between 40 and 100 Ω?
Yes
Between 40 a
nd 100 Ω! THREE (3) possible bad components (see COMPONENT
LIST below). One at a time, swap in each component and in between perform the
DiaSafe
®
test until the syringe plunger is drawn indicating the last component
swapped in was the problem.
COMPONENT LIST:
1)
Actuator-Test Board;
2)
Actuator cable;
3)
Distribution
board.
No
Less than 40 Ω OR more than 100 Ω. TWO (2) possible bad components:
1)
Bad
valve #28 blue wire harness OR;
2)
Bad valve #28.
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...