2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 326
CO- 1.0.5.2 CHAMBER #69 SAYS “AIR”
:
A)
Figure below
,
inside the distribution board, locate the Air Sensor’s female connector!
4
th
connector cap i.e. 5
th
position
FROM THE LEFT. If CBE modified it plugs into the CBE board
higher than the others!
Continue to part B next page
LEFT BLANK INTENTIONALLY