CO- 8.2.0 ISOLATE VERY INTERMITTENT ACID PUMP CYCLES
Call debug screen 0. Figure (right), WITHOUT LOOKING AWAY, watch
Acid
versus
AAcid
for one (1) minute. Are they remaining always /- 3 of each other?
Yes
AAcid
=
Acid!
See procedure number
(page 368).
No
AAcid
is NOT remaining within 3 of
Acid!
Proceed to
page 397
...
CO- 8.2.1 AACID = ACID / ISOLATE BICARB PUMP CYCLES
Figure right, watch
Bic
versus
ABic
for one (1) minute. Are they remaining always
/- 3 of each other?
Yes
ABic
=
Bic!
(page 368).
No.
ABic
is NOT remaining /- 3 of
Bic!
Figure (right), watch the
Bicarbonate (BIC) pump’s ‘dot’ for THREE (3) minutes. It should REMAIN
blue. TWO (2) possible scenarios:
1)
IF (and ONLY if) the ‘dot’ is EVER white:
Proceed to
page 374
, procedure number
2) IF the ‘dot’ REMAINS BLUE:
Proceed to
page 397
CO- 8.2.2 ABIC = BIC
a) If not already, allow the external flow indicator’s ‘bob’ to begin moving up and down
b) Based on the external meter’s reading versus TCD, TWO (2) possible scenarios:
1) IF (and ONLY if) LESS THAN TCD - (minus) 0.4:
(page 369).
2) IF MORE THAN TCD + 0.4 i.e.:
page 369
).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...