2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 307
T- 7.2.6 ‘BOB’ WAS NOT RISING BUT IS NOW / RESET INTERNAL ALARM LIMIT
a)
IMPORTANT!
RETURN NTC #3’s connector to
2
nd
distribution board position from the left “MON-NTC”.
b) Select the
[TEMP DAC]
data box, it turns bright yellow.
c) Set
[TEMP DAC]
to “130” then ‘sharply’ press
‘CONFIRM’ ONCE.
d)
IMPORTANT!
Allow five (5) minutes BEFORE continuing
to part e!
e) ‘Sharply’ press ‘CONFIRM’ TWICE. The
[Monitor Reference]
data box turns pale yellow / white.
f)
DO NOT press ‘CONFIRM’ until instructed! Select [Monitor Reference]
it turns bright yellow.
g) Ignoring the external meter adjust
[Monitor Reference]
to 34.0° C.
h) ‘Sharply’ press ‘CONFIRM’ TWICE to turn
[Monitor Reference]
GRAY! Does the GRAY
[Monitor
Reference]
box remain less than 35.0° C?
Yes
[Monitor Reference]
remains less than 35.0° C! Press and release ‘CONFIRM’ key until
the calibration is saved then return to (ABOVE) procedure number
page 305
No
[Monitor Reference]
does NOT remain less than 35.0°C!
Read before performing!
Swap the
listed components (see COMPONENT LIST below) one at a time, with known good, and in
between return to (ABOVE) procedure number
page 305
) and test each new
component until the
[Monitor Reference]
data box remains less than 35.0° C.
COMPONENT LIST:
1)
NTC #3*;
2 )
Actuator-Test Board
**
;
3)
Sensor Board
**;
4)
Sensor
Board cable;
5)
Functional Board (possibly IC20).
*
(page 218)
**
To LOCATE these boards refer to
(page 9)
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...