2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 634
P- 1.1.0 ISOLATE POWER LOGIC BOARD CABLE (PIN 4)
a) Turn the machine off and
UNPLUG
it.
CAUTION!
Electrocution hazard if not unplugged!
b) Slide the power supply away from the cabinet to see
the Power Control Board inside.
c) ENSURE ALL cables are plugged in securely.
If
not, this may be the problem!
d) Figure right, un-mount the board from its white
plastic clips.
e) Remove the two
front
screws and lay the rear
power supplies panel down to access to the rear
(solder) side of the X2 cable.
f)
IMPORTANT!
Plug the machine in!
CAUTION!
High voltage now present!
g)
below
, measure at the rear (solder) side of the Power Control board’s X2
connector, at
pin 4
(BOTTOM row, SECOND pin from the left). More than 10.0 volts DC?
Yes
More than 10.0 volts! TWO (2) possible bad components:
1)
20 pin Power Logic ribbon
cable (X2) OR;
2)
Power Control board.
No Less than 10 volts! See procedure number
(page 635).
Figure 126 – Power Control Board Solder Side