2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 293
T- 4.4.0 CHECK TEMP DISPLAY CALIBRATION (1)
a)
Place the machine into Service Mode → Options → Hardware Options.
b) Does the ‘Yes’ box at ‘
Temp Comp
’ have a blue ‘X’ in it?
Yes
Place the ‘X’ in the “No” box and press ‘CONFIRM’ (the ‘X’ turns
blue). Post a note that ‘
Temp Comp
’ is off then see procedure
No Temp Comp = No! See procedure number
(page 293).
T- 4.4.1 CHECK TEMP DISPLAY CALIBRATION (2)
a) From the Service screen menu
→ Calibrate Sensors → Temp Control.
b) Press ‘CONFIRM’ TWICE. The screen’s
[TEMP DAC]
data box turns pale yellow / white.
c) Select
[TEMP DAC].
It turns bright yellow and set its value to “20”
d) Press ‘CONFIRM’.
[TEMP DAC]
returns to pale yellow / white.
e)
Set your volt meter to AC voltage (~ V, V
AC
)!
f)
IMPORTANT!
Allow
the screen’s
[Monitor Reference]
data box drop to 39.0° C or less.
g) Measure again at the distribution board’s Heater Connector between the BLUE and BROWN wires.
More than 100.0 volts AC now?
Yes
More than 100.0 volts AC! See procedure number
(page 294).
No Less than 10 volts! See parts a THROUGH c below:
a) Select the
[TEMP DAC]
window
and set it to “130”!
b) Save the calibration.
c) Proceed to
page 302
, procedure number
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...