2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 108
F- 12.2.0 NEGATIVE 250 MMHG WAS ACHIEVED / PRESSURE TEST
Does meter pressure HOLD, +/- 15 mmHg, for one (1) minute?
Yes
Pressure holds! See procedure number
(page 108).
No Pressure DOES NOT hold! Proceed to
page 109,
F- 12.2.2 PRESSURE HOLDS / VERIFY DIALYSATE PRESSURE
a) ENSURING the external meter is HOLDING negative (-)250 +/- 5 mmHg press ‘CONFIRM’ twice to
save the calibration.
b) Figure right, TWO (2) possible scenarios based
on if an “Operator Error” banner occurs:
1)
IF (and ONLY if) “Operator Error” occurred:
page 110
2) IF “Operator Error” did NOT occur:
See parts a THROUGH e below:
a) Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program (
“Select Program” → ‘Dialysis’ →
‘CONFIRM’)!
c) From the Home screen, set [Dialysate Flow] to 500 ml/min.
d)
IMPORTANT!
Allow five (5) minutes BEFORE continuing to part e
e) Return to
page 105
, procedure number
LEFT BLANK INTENTIONALLY