2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 550
OVER- 7.0.0 ISOLATE VALVE #41
Figure right, unplug valve #41 from distribution board
position #27 (“V27, “IN-V”). TWO (2) possible scenarios:
1)
IF massive overflow STOPS:
See procedure
(page 550).
2)
IF massive overflow does NOT stop:
ENSURING
Valve #41 was unplugged,
perform parts AND b
below:
a) Turn the water OFF! Referring
(page 193), trace valve #41’s wiring harness from the
distribution board to ENSURE it terminates at valve #41’s solenoid terminals.
b) If the harness terminates at valve #41, valve #41 is bad (to LOCATE refer to
(page 193)).
OVER- 7.0.1 ISOLATE FLOAT SWITCH
a) With the machine AND water ON, CAREFULLY return valve #41 to distribution board position “IN-V”. If
(and ONLY if) ‘massive’ overflow continues see part b.
b) Figure right, unplug the float’s connector from distribution
board position “X5, FLOAT-SW”. Does ‘massive’ overflow
continue?
Yes
Overflow continues! ENSURING the float was
unplugged, the Actuator-Test Board is bad (see
No a) Trace the float’s wiring harness from the distribution board to ENSURE it terminates at the
float. NOTE: To LOCATE the Float refer to
(page 129).
b) If the harness terminates correctly replace the float.
c) With the machine AND water ON, if (and ONLY if) massive overflow continues see part d.
d)
IMPORTANT!
Turn the machine OFF to prevent
electrocution then see part e.
e) Figure right, remove the heater and check its port
into the top of the Hydrochamber for heat damage
(i.e. melting). If (and ONLY if) damage is located it is
necessary to replace the Hydrochamber. If (and
ONLY if) heat damage is NOT located see part e.
e) Reinstall the heater.
f) With the machine AND water ON, if (and ONLY if)
massive overflow continues see part g
g) FOUR possible bad components:
1)
Bad actuator cable OR;
2)
Bad Sensor Board cable OR;
3)
Bad distribution board OR; 4
)
Bad motherboard.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...