2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 48
F- 5.1.0 BALANCING CHAMBER VALVES WIRED CORRECTLY
a) Call debug screen 6.
b) Watch
BC Switch
(middle column)
for thirty (30) FULL seconds. Does it REMAIN
ALWAYS = 897?
Yes
BC Switch
ALWAYS = 897! See procedure number
(page 48).
No
BC Switch
is NOT always = 897!
Proceed to
page 49
, procedure number
F- 5.1.1 BC SWITCH ALWAYS = 897 / ISOLATE PRESSURE SENSOR CONNECTIONS
a)
Figure below
,
VERY CAREFULLY
trace BOTH wire harnesses from distribution board
connectors “X9, P-Dial” AND “X10, CFS” to ENSURE they go to the correct (blue) Pressure
Transducer.
b) Are BOTH Pressure Transducers connected properly?
Yes
BOTH are connected properly! See procedure number
(page 49).
No
Connect them properly i.e. CFS #10 → “X10, CFS”; Dialysate Pressure #9 → “X9, P-Dial”
and allow one (1) minute. Call debug screen 1 to watch
Flow Error
for four (4) minutes. If it
remains = 0 this may have fixed the problem. If it EVER = 1, even just once, continue
procedure number
Figure 10 – Pressure Sensors
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...