2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 473
TMP- 9.0.0 VALVE #28 LEAKING / ISOLATE ACTUATOR-TEST BOARD
DiaSafe
®
filter ‘Test Valve’ #28 is leaking:
a) Figures right and below, trace the wires
from Valve #28’s distribution board
position (“V28, SAMPLE”) to ensure it
connects to the correct valve.
b) Unplug Valve #28 from distribution board
position “V28, SAMPLE”. Does the Valve
continue to leak?
Yes
Valve #28 continues to leak. Valve #28 is bad!
No Valve #28 stops leaking! Turn the machine off and replace the Actuator-Test Board.