2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 122
F- 16.0.0 LESS THAN 150 ML COLLECTED / ISOLATE RESTRICTOR #48
This procedure attempts to clear an air lock. First Restrictor #48 is checked to see if it is plugged.
a) Figure right, a
60 ml syringe
*
, with its plunger pushed ALL
THE WAY IN is required!
*
Using a smaller syringe may cause error!
b) Attach the syringe, to the DEAERATION PUMP’s clear (INPUT) tubing removed earlier.
c) PULL on the plunger to the end of the barrel! If Restrictor #48 is plugged you will feel very strong
resistance and may not be able to pull the plunger all the way out. Very strong resistance (Yes or No)?
Yes
Strong resistance! Restrictor #48 may be plugged with debris
1
. Replace the tubing segment
that contains Restrictor #48*. *To LOCATE Restrictor #48 refer to
(page 120)!
1
Debris indicates FIVE (5) possibilities:
1)
Inadequate acid cleaning (white bicarbonate
precipitate) OR;
2)
Excessive O-ring lubrication OR;
3)
Inadequately filtered incoming
water OR;
4)
Degrading Deaeration Pump Head OR;
5)
Degrading Heater element.
No Very little resistance! Restrictor #48 is okay! See part A below:
A) Remove the syringe and, from the tubing, MEASURE flow for one (1) minute. More than 130
ml collected?
Yes
More than 130 ml! See procedure number
F- 16.0.1
(page 122).
No Less than 130 ml! See part B.
B)
Push the plunger
b
ack into the syringe barrel then
reattach the syringe to the
Deaeration Pump’s input tubing.
C) Pull the plunger to the end of the barrel!
D) Repeat parts A through D up to
TEN (10) TIMES
OR until if more than 130 ml every minute
is MEASURED?
Yes
More than 130 ml every minute! See procedure number
F- 16.0.1
(page 122).
No After ten (10) attempts you CANNOT achieve more than 130 ml every minute!
Proceed to
page 123
, procedure number
F- 16.0.1 MORE THAN 130 ML COLLECTED / DOES IT CONTINUE?
Direct the tubing into the bucket for up to FIVE (5) FULL minutes. Flow will either:
1)
Continue at more than
130 ml per minute OR
2)
Stop. TWO (2) possible scenarios:
1)
IF (and ONLY if) flow continues at more than 130 ml every minute:
The air lock has been
eliminated for now! Proceed to
page 125
, procedure number
2) IF flow stops:
Proceed to
page 123
, procedure number
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...