2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 61
F- 6.5.0 TMP ALARM?
a) From the Home screen, set [Dialysate Flow] to 500 ml/min and press CONFIRM!
b) Allow thirty (30) seconds THEN TWO (2) possible TMP window scenarios:
1)
IF the TMP window is NOT red:
See
(page 61).
2) IF the window is RED (TMP alarm): RESET SEQUENCE: A)
Press and release the ‘Reset’ key
then press and hold it for three (3) seconds;
B)
Allow thirty (30) seconds;
C)
If a TMP alarm
reoccurs attempt the RESET SEQUENCE up to twice more BEFORE continuing to procedure
number
(page 61).
F- 6.5.2 ISOLATE TMP STABILITY
TMP is STABLE if its window remains white AND does NOT change more than 60 in three (3) minutes.
TWO (2) possible scenarios:
1)
IF (and ONLY if) TMP is UNSTABLE:
(page 61).
2) IF TMP is STABLE:
A procedure, in a different Section, is performed next
.
NOTE
this page and
procedure number because you may be prompted to return to here. Perform parts A through C below:
A) BEFORE continuing to part b, proceed to
page 535
, to perform
A LEAKING BALANCING CHAMBER DIAPHRAGM
B) If a balancing chamber diaphragm leak was not located in part a, return to Dialysis Program
(
“Select Program→ ‘Dialysis’ →
‘
CONFIRM’)!
C) Proceed to
page 62
, procedure number
F- 6.6.0 TMP IS UNSTABLE
a) A procedure, in a different Section, is performed next.
NOTE
this page and procedure number (F- 6.6.0)
as you may prompted to return to here.
b) BEFORE continuing to part c, proceed to
page 527
, to perform
c) If a leaking balancing chamber valve was not located in part b, return to Dialysis Program (“Select
Program”
→ ‘Dialysis’ →
‘
CONFIRM’)!
d) ENSURING Dialysate Flow is ON (Flow on/off lamp NOT blinking) call debug screen 6.
e) WITHOUT LOOKING AWAY, watch
BC Switch
for up to five (5) minutes. ENSURING a “No Water”
alarm NEVER occurs does
BC Switch
EVER = 897 or more even just once?
Yes
BC Switch
= 897 or more (constant or once in a while)! Proceed to
page 62
, procedure
No
BC Switch NEVER, EVER = 897 or more!
If TMP remains unstable proceed to
page 440
. If TMP is stable the flow problem is no longer presenting.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...