2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 605
Yes
BOTH more than 23.0 volts! Proceed to page
, procedure number
No BOTH are less than 23.0 volts! TWO (2) possible scenarios:
1)
IF (and ONLY if) BOTH are less than 23.0 volts:
2) IF AT LEAST one is more than 23.0 volts:
The Power Logic X2 ribbon cable may be bad.
P- F.5.0 I24V-A
a) Spread the card cage side panels open then gently drop the front panel down to access the
motherboard’s nine (9) pin TEST Connector.
b) ENSURE the machine is on (fan running)!
c) ENSURE the meter’s black lead REMAINS connected to chassis ground!
d)
Per the Figure below
, measure from the TEST Connector’s
pin 7
(three (3) pins from the right).
More than 23.0 volts DC (Yes or No)?
Yes
More than 23.0 volts DC! TWO (2) possible bad components:
1)
Power logic cable OR;
2)
Bad thermal (BROWN) fuse S19
*
(5 Amp).
*
To LOCATE fuse S19 refer to
(page 651)! With the machine off, it can be
checked by measuring RESISTANCE (Ω) BETWEEN its two terminals. A good fuse
measures less than 0.5 Ω!
No
Less than 23.0 volts DC!
(page 606).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...