
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 211
VEC- 1.5.0 VALVE #24 OKAY / ISOLATE VALVE #25
a) With the shunt door open, the dialyzer connector NOT in the shunt AND still in Heat Disinfect OR Rinse,
ENSURE the external flow indicator’s ‘bob’ is rising!
b) Return the dialyzer connector to the shunt but
DO NOT
close the door!
c) Watching for thirty (30) seconds, does the external flow indicator’s ‘bob’ rise at least ½ way up now?
Yes
Bob moving! Something changed!
A)
Turn the machine OFF;
B)
Close the door;
C)
Turn
the machine on and return to Heat Disinfect OR Rinse;
D)
If a Flow Error reoccurs return to
(page 159).
No ‘Bob’ not moving!
NOTE
ONLY
VALVE #25
will be checked and proceed to
page 192
LEFT BLANK INTENTIONALLY
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...