2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 458
Yes
-250 mmHg holds!
A)
IMPORTANT!
Remove the clamps from both Four-Way tubing
segments then;
B)
Leaving [Dialysate Flow] off, proceed to
page 466
, procedure number
No -250 could not be achieved or did NOT hold! Either the transducer protector is wet OR a Four-
Way assembly tubing connection is leaking! See parts A and B below:
A)
Locate and repair the leak. Consider using another Four-Way assembly!
B)
Return to (ABOVE) procedure number
(page 456).
LEFT BLANK INTENTIONALLY