2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 190
CLEAN- 7.3.0 ISOLATE VALVE CIRCUITS
a) Figures below, you will check
four (4) Valves
:
1)
Valve #24;
2)
Valve #38;
3)
Valve #39;
4)
Valve
#43.
b) THREE (3) checks on EACH Valve! Trace the blue wires from between the distribution board to their
solenoid terminals:
CHECK #1
ENSURE no ‘pinched’ or damaged wire insulation
CHECK #2
ENSURE the wires terminate properly at the solenoid terminals
CHECK #3
ENSURE the terminals show no signs of ‘green’ corrosion
c) If a problem was found (and repaired) this may have solved the problem. If a problem WAS NOT
(page 191).