2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 355
STEP #3:
Watch the external meter for thirty (30) seconds. Its reading MUST BE stable i.e. does NOT
change more than +/- 0
.
15 mS?
Yes
Conductivity is stable! Continue to STEP #4.
No a) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ →
‘
CONFIRM’)
b) See (ABOVE) procedure number
(page 350).
STEP #4:
Is the screen’s
Frequency
window between ten-thousand (10000) and twelve-thousand
(12000) Hz (Hertz)?
Yes
Frequency
between 10000 and 120000 Hz! Continue to STEP #5.
No
Frequency
is NOT between 10000 and 12000 Hz! See parts a AND b below:
a) Figure below, trace Cond Cell #7’s cable to ENSURE it plugs into the
Distribution board PROPERLY.
If not this is the problem!
b) FIVE (5) possible bad components (see Component List below). Swap in, one
at a time with known good and in between repeat (ABOVE) procedure number
(page 354) to test each.
COMPONENT LIST:
1)
Conductivity Cell #7;
2)
Sensor Board cable;
3)
Functional Board;
4)
Distribution board;
5)
Motherboard
STEP #5:
If Cond Cell #7’s reading stabilizes,
“5. Conductivity value is stable….”
appears and the
[Conductivity]
data box turns pale yellow / white?
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...