2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 92
F- 9.0.0 ACFS BETWEEN 3.1 AND 8.0 / ISOLATE FLOW PUMP
a) Return BOTH concentrate connectors to their rinse ports!
b) This procedure requires a psi gauge.
ENSURE
it reads 0 psi before installing it!
below, tee the gauge into the Flow Pump’s WHITE (output) tubing.
d)
Tie wrap both sides of the gauge tubing to prevent leaks and false readings!
e)
Place the machine in RINSE!
f) ENSURING a “No Water” NEVER appears, call debug screen 0. If debug does not appear press
‘Escape’ then call screen 0.
g) Allow Valve #43’s ‘dot’ (Figure right) to turn blue then WHITE again! While
white, does gauge pressure CYCLE, about every three (3) seconds, to
between
35 and 36 psi?
Yes
Between 35 and 36 psi! Proceed to
page 101
, procedure number
No
IS NOT between 35 and 36 psi! ENSURING the machine was in RINSE AND no leaks, TWO
(2) possible scenarios:
1)
IF (and ONLY if) pressure is too low:
DO NOT calibrate instead proceed
page 93
, procedure number
2) IF pressure is too high:
A)
Per the Figure below, adjust Valve #78 until pressure cycles
to between 35 and 36 psi THEN;
B)
Proceed to
page 101
, procedure number
Figure 21 –
Hydraulics, Rear View
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...