2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 433
FIL- 5.1.0 FILACT = 0 / ISOLATE VENOUS PRESSURE (VEN)
a)
Turn Dialysate Flow OFF i.e. Flow on/off lamp blinks!
b) Call debug screen 0. ENSURE all eight balancing chamber valves (#31 through #38) ‘dots’ are
REMAINING white (Flow is Off)!
c)
IMPORTANT! Dialysate Flow REMAINS OFF till instructed!
d)
Per the Figure below
, attach a syringe with a piece of tubing attached that will fit SNUG to the
Level Detector’s INNER P
ven
port.
e) Call debug screen 1. Push on the syringe plunger until
VEN
(middle column) is
between 400 and 450.
f) Clamp the syringe tubing to HOLD the pressure!
g) If
VEN
falls more than 3 mmHg in 30 seconds there is a leak at the P
ven
port OR inside the module.
h) Leaving the clamp in place to keep
VEN
at more than 400, see procedure number
(page
LEFT BLANK INTENTIONALLY