2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 26
F- 1.0.5 RESET MACHINE / DEFEAT IDLE FLOW
a)
Turn the machine OFF!
b)
Open the shunt door and leave it OPEN till instructed!
c) Turn the machine on and re
turn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
d) From the Home screen, select the [Dialysate Flow] window.
e) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
f) Per the Figure below, check the Flow AND Deaeration Motor shafts for counterclockwise (CCW)
rotation. BOTH rotating CCW?
Yes BOTH motors rotating! See procedure number
No A motor is NOT rotating CCW! Perform parts a THROUGH c below:
a)
Close the shunt door!
b)
Call debug screen 1. ONLY if
FILACT
always = 0 see part c. If
FILACT
= 1 return to
(page 24).
c)
NOTING
which motor is not rotating properly, proceed to
page 130
Figure 7 – Deaeration AND Flow Motor