2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 167
CLEAN- 2.0.2 MORE THAN 130 ML
Direct the tubing into a bucket for up to FIVE (5) FULL minutes. Flow will either:
1)
Continue at more than
130 ml per minute OR
2)
Stop. TWO (2) possible scenarios:
1)
IF (and ONLY if) flow continues at 130 ml per minute:
2) IF flow STOPS:
Proceed to
page 169
, procedure number
CLEAN- 2.0.3 FLOW CONTINUES AT MORE THAN 130 ML PER MINUTE / ISOLATE VALVE #39
Cleaning Programs open Valve #39. This procedure isolates Valve #39:
a) Per the Figure above, at the bottom of the Hydrochamber, clamp the tubing segment that contains
Restrictor #48 at the LOCATION SHOWN!
b) Direct the Deaeration Pump’s clear tubing into the
1000 ml
cylinder.
c)
If the Flow Error was occurring in Heat Disinfect return to it! If not place the
machine in RINSE!
d) Measure for ONE (1) MINUTE! More than
two hundred (200) ml
collected?