2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 614
P- G.5.5 SYMPTOM NOT PRESENT / ISOLATE SENSORS AND PUMPS
a)
IMPORTANT! To prevent damage turn the machine OFF!
b)
IMPORTANT!
Reconnect the Sensor Board cable!
c) One of 13 Sensors (NTC #2 - NTC #44) or one of five (5) Pumps (#16 - #22) may be causing the
problem. This procedure locates the bad component.
Figure 118 – Distribution Board
-
d) With the machine off, per the Figure above, your choice, unplug ONE of the Sensors or ONE of the
Pumps from the Distribution board.
e)
ENSURE the meter’s black lead remains connected to chassis ground!
f)
IMPORTANT! Turn the machine ON (fan running)!
g) IGNORE grinding and / or UF and / or acid and / or bicarb pump alarms and / or Flow Errors.
h) Re-m12 and / or +5 volts (within range now?). If the screen is on select the ‘Dialysis’ button but
DO NOT
press ‘CONFIRM’ THEN call debug screen 1. If the symptom was intermittent allowing
sufficient time, does the symptom you were troubleshooting reoccur?
Yes
Symptom reoccurs! Repeat parts d through h until ALL 13 Sensors (including Blood Leak) AND
ALL 5 pumps have been unplugged OR until the symptom does NOT reoccur indicating the last
component plugged in or its cable is the problem.
NOTE:
If after ALL components have been unplugged AND the symptom reoccurs, turn the
machine off and replace the Sensor Board cable with a known good then repeat parts e
through h. If the symptom reoccurs see procedure number
(page 615).
No Screen on and / or good voltages! The last component unplugged or its cable may be causing the
problem.
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...