DVF- 4.0.0 ISOLATE VALVES #24 AND #43 SOLENOID CIRCUITS (1)
a)
IMPORTANT! Turn the machine OFF!
b) Per the Figure below, THREE (3) checks while tracing Valve #43’s AND #24’s wires from the distribution
board to BOTH valves inside the hydraulics.
CHECK #1:
ENSURE no ‘pinched’ or damaged wires!
CHECK #2:
Figure right,
ENSURE the wires terminate PROPERLY at
the valve’s solenoid terminals)!
CHECK #3:
ENSURE the terminals show no signs of ‘green’ corrosion!
(page 657).
DVF- 4.1.0 ISOLATE VALVES #24 AND VALVE #43 SOLENOID CIRCUITS (2)
a) Figure right, open Valve #24’s AND #43’s female distribution
board connector caps.
b)
The wires MUST connect between the TOP and
BOTTOM terminals! If NOT, this is the problem!
c)
IMPORTANT! Set your CALIBRATED volt meter
to RESISTANCE (Ω)!
d) Where a blue wires are attached, place one meter lead on one of
the terminals and the other lead on the other terminal.
e) Figure right, reading the meter’s numeric AND units
display! TWO (2) possible scenarios next page:
1)
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...