2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 447
TMP- 2.0.1 POSITIVE PRESSURE LEAK CHECKS
Small external leaks cause BIG problems! For the following USE A FLASHLIGHT! FEEL AROUND!
NOTE:
Not being THROUGH may cost unnecessary troubleshooting time!
a) THREE (3) leak checks:
CHECK #1:
Figure right, the ENTIRE length of the
dialyzer lines ESPECIALLY the quick
connectors!
CHECK #2:
The ENTIRE hydraulic compartment!
CHECK #3:
Figure right, DiaSafe
®
filter housing and tubing connections.
b) Figure below, any leak seen through Diasafe
®
Filter #92*?
*
Water staying inside the filter is NORMAL!
Yes
Valve #28 is leaking! Proceed to
page 473
, procedure number
No leaks seen! See procedure number
(page 448)!
Figure 68 – DiaSafe
®
Filter / Test Valve #28 / Filter #92
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...