2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 587
DIASAFE- 4.0.0 DIASAFE TEST FAILED / ISOLATE ATMP AND FILLING PROGRAM
A)
RESET ALL alarms
!
B) Turn the Blood Pump on so that it is rotating at more than 100 ml/min.
C) Call debug screen 1.
ATMP
AND
FILACT
will be
watched below. If
FILACT
EVER = 1 may indicate a hydraulic air leak causing
ATMP
to be unstable or not reach its specified goal of 300!
D) To avoid error read parts E and F BEFORE performing them.
E) To turn UF on, press the front panel’s UF on/off key. The UF lamp (Figure
right) MUST STAY ON SOLID!
F) Without looking away, watch
FILACT
AND
ATMP
. If
ATMP
reaches 300,
within forty (40) seconds, turn UF off
IMMEDIATELY
i.e. UF lamp OFF!
TWO (2) possible scenarios:
1)
IF ATMP reached 300 within 40 seconds:
2) IF ATMP DOES NOT reach 300 within 40 seconds:
Based on
FILACT
, TWO (2) possible
scenarios below:
Scenario #1:
IF FILACT WAS EVER = 1:
A)
Turn UF OFF then!
B)
Allow two (2) minutes
then;
C)
Repeat parts E and F of procedure number DIASAFE- 4.0.0 then;
D)
If (and ONLY if)
FILACT
= 1 reoccurs turn UF OFF and proceed to
page
SECTION 8 – FILLING PROGRAM PROBLEMS
Scenario #2: IF FILACT was ALWAYS = 0: A)
To check the UF Pump, perform the UF
Pump Calibration per the 2008 Calibration Procedures booklet then;
B)
If (and
ONLY if) UF volume is okay perform the Dialysate Pressure Calibration per the
2008 Calibration Procedures booklet
.
DIASAFE- 4.0.2 ATMP REACHED 300
Watching
FILACT
AND
ATMP
,
allowing ten (10) seconds for stabilization,
ATMP
should stay more than
300 for twenty (20) seconds. TWO (2) possible scenarios below.
1)
IF ATMP stays 300 OR more:
(page 588).
2)
IF ATMP DOES NOT stay more than 300:
Based on
FILACT
, TWO (2) possible scenarios:
Scenario #1:
IF FILACT was EVER = 1: A)
Allow two (2) minutes then;
B)
Repeat procedure
number DIASAFE- 4.0.0 (page 587);
C)
If (and ONLY if)
FILACT
= 1 reoccurs
proceed to
page 409
SECTION 8 – FILLING PROGRAM PROBLEMS
Scenario #2:
IF FILACT was ALWAYS = 0:
ENSURING UF is off proceed to
page
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...