2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 434
FIL- 5.2.0 VEN MORE THAN 400
a) Figure right, connect the Four-Way Assembly
(P/N 150034) to the dialyzer connectors.
b) Place the Four-Way at dialyzer level!
ENSURE a transducer protector IS NOT in
the ‘to syringe’ tubing segment!
c) If using a
NEO-2
attach to the +Port (top, red
port). If using a
90XL
attach to the Pressure
Module’s Gauge Port.
d) DO NOT allow tension in the Four-Way tubing
segments!
e)
IMPORTANT! CLOSE THE SHUNT
DOOR!
f) Call debug screen 2. To ENSURE the shunt door is closed
CVRCLS
(2
nd
column from left)
=
1!
g) PULL on the syringe plunger. Can you achieve approximately negative (-)250 mmHg on the external
meter?
Yes
-250 mmHg achieved! Clamp the four ways’s ‘to syringe’ tubing segment to keep the
pressure then see procedure number
(page 434).
No ENSURE the transducer protector, at the meter, is not wet OR consider replacing it. If OKAY,
proceed to
page 435
, procedure number
FIL- 5.3.0 -250 ACHIEVED / PRESSURE TEST HYDRAULICS (2)
Watch for one (1) FULL minute. Does pressure HOLD +/- 15 mmHg?
Yes
Pressure holds! Proceed to
page 436
, procedure number
No Pressure DOES NOT hold! See procedure number
(page 435).
Содержание 2008K2
Страница 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Страница 197: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 193 Figure 35 Hydraulics Top View Figure 36 Hydraulics Rear View ...
Страница 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Страница 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Страница 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...